Imaging and Obese Patients
|
|
- Obesity can limit radiographic evaluation
- Image quality is degraded by a patient's large body habitus; Ultrasound is most affected
- CT, MRI, and X-ray have table weight limits that cannot be exceeded
- The bore diameter of CT (70cm) and MRI (60cm) scanners is too small for the body girth of some
patients
|
Current
estimates indicate that 4.7% of the American population today is
extremely obese, with a body mass index (BMI) >40; an increase from
0.8% in 1960 and 2.9% in 1988. Obesity presents many challenges in
health care. In addition to increasing risk for the development of
chronic disease, obesity is a limiting factor in radiology used in all
aspects of medical diagnosis and treatment. These limitations include
both logistical and image quality issues. For example, obese patients
need larger and stronger wheelchairs and gurneys for transportation,
while physical limitations of the imaging equipment in terms of table
strength and bore size may make the desired examination impossible.
Weight limits for examination tables are set by the manufacturers and
cannot be exceeded without endangering both the patient and the
equipment. Large body habitus also degrades image quality, making it
difficult or impossible to obtain adequate images for clinical
interpretation.
Factors that Limit Imaging
Ultrasound
Ultrasound images are affected by fat to a greater degree than any
other imaging modality. Ultrasound energy is attenuated by fat tissue,
especially at higher frequencies. For example, at the frequency range
(3-7 MHz) normally used for abdominal imaging, 50% of the beam
intensity (watts/cm2)
is attenuated when the beam travels through 1 cm of fat and the sound
level drops by 3 decibels (dB). In an obese patient with 8 cm of
extra-peritoneal fat, there would be about 6% of the original beam
intensity or a 24 dB drop in sound level before the beam enters the
peritoneal cavity. Attenuation is less at lower frequencies but results
in decreased image resolution.
Plain Film Radiography and Fluoroscopy
Image quality on plain radiographs is also severely limited by
attenuation. Increasing the current and exposure time can improve
image quality but also increases the radiation dose to the patient and
cause motion artifact due to increased exposure times. Some obese
patients are too large to image using a single 14 x 17 inch film
cassette and require multiple cassettes. When they exceed the
weight limit for the x-ray tables (480 lbs), patients can sometimes be
imaged while standing.
X-ray fluoroscopy is also restricted to patients who do not exceed the
table weight limit, which in this case is usually 350 lbs. In addition,
the clearance between the table and the fluoroscopy tower is 45 cm.
|
|
|
Percentage of All Imaging Studies
“Limited by Body Habitus”
|
| Chest x-ray |
0.5% |
| Abdominal x-ray |
0.25% |
| Ultrasound |
2% |
| Chest CT |
0.25% |
| Abdominal CT |
0.4% |
| MRI |
0.1% |
CT
X-ray attenuation also limits image quality for CT images, although to
some degree attenuation can be compensated for by increasing the tube
current and decreasing the rotation speed of the gantry. However, there
are absolute limits on patient’s weight and girth. The weight limit for
the CT tables is 425-450 1bs, depending on the scanner. The bore of the
gantry has a diameter of 70 cm, which can sometimes be too small for
patients who do not exceed the weight limit. Although it is sometimes
suggested that veterinary CT scanners are an option, this is rumor
only. The scanners used are the same as those used clinically and
we know of no veterinary hospitals capable of imaging obese
patients.
Nuclear Imaging
Image
quality in nuclear medicine severely falls off in obese patients due to
absorption and photon scattering, which is greater for lower energy
isotopes. Therefore, 99mTc, which emits higher energy photons than 201T-thallium,
is likely to provide higher quality images in obese patients. Low image
quality is especially problematic for myocardial perfusion SPECT in
obese patients, since the low number of photons detected makes it
difficult to know whether to attribute a drop in photons detected to a
soft tissue attenuation artifact or myocardial ischemia.
The quality of PET images is also lower in obese patients for two
reasons. First, image quality is degraded due to the high scatter in
soft tissues and increased photon attenuation in a larger body mass.
Secondly, it is not possible to give a large enough dose of isotope, as
the maximum dose recommended for patient safety can often be well below
the mCi/kg dose required for optimal image quality.
MRI
MRI
image quality is least affected by obesity although increased body
habitus introduces noise and the large field of view needed decreases
the in-plane resolution of the images. The main limitations of MRI are
the size of the bore and the table weight limits, which prevent imaging
of large patients. MRI bore diameters are smaller than those of CT
scanners and contact with the bore creates eddy currents that degrade
MR images. Open bore MRI systems, which are not available at MGH, can
be used to image larger patients, but they generally offer lower field
strength, resulting in degraded image quality, longer exam times
causing patient discomfort and motion artifact on images, and inability
to perform exams requiring dynamic intravenous contrast administration
(MRA, liver scans, etc.).
|
|
Weight and Size Limits for Imaging and Radiologic Intervention
|
|
Maximum Weight |
Bore Diameter or
Maximum Distance |
Field of View |
Radiography
Prone
Standing |
480 lbs
None |
N/A |
14 x 17 in.
14 x 17 in. |
| Ultrasound |
None |
N/A |
N/A |
CT
4-16 slice multi-detector scanners
64 slice multi-detector scanner
(Main campus) |
425 lbs
450 lbs |
70 cm
70 cm |
50 cm
50 cm |
MRI
Siemens (Main campus and Yawkey)
GE (Main campus and off-campus imaging centers) |
440 lbs
325 lbs |
60 cm
60 cm |
50 cm
45-50 cm |
| Nuclear Medicine |
400 lbs
|
|
|
Fluoroscopy
Other
Innova (Yawkey only) |
350 lbs
450 lbs |
45 cm |
N/A
|
Interventional Procedures
Angiography
Spine
Mammography biopsy
Other
|
350 lbs
350 lbs
350 lbs
350-400 lbs |
N/A
|
N/A
|
| Note: 60 cm diameter = 74 inches circumference; 70 diameter = 87 inches circumference |
 |
| Chest x-rays of normal weight (left) and obese (right) patients. The latter image quality is limited by body habitus |
|
 |
| Ultrasound
images of liver in normal weight (left) and obese (right) patients. The
latter image quality is limited by body habitus |
 |
| Abdominal CT images of a normal weight (left) and obese (right) patient. Note image artifact due to patient contact with bore |
|
Impact on Patient Care
If
patients cannot be imaged by the modality of choice, this can adversely
affect their care. This is particularly important for patients
who have undergone weight loss surgery. For example, all patients who
have laparoscopic gastric bypass surgery at MGH routinely undergo a
follow-up Gastrografin® swallow to check for an anastomotic
leak. However, if patients exceed the size limit for fluoroscopy,
evaluation for a leak is limited to serial portable chest and abdominal
x-rays, which are significantly less sensitive. Imaging of other
complications of bariatric surgery, such as abdominal abscess or
collection, is nearly impossible if the patient cannot undergo
CT. Diagnosis of other post-operative complications such as deep
vein thrombosis (with ultrasound or pelvic CT) and pulmonary embolism
(with chest CT angiography) is also compromised.
Difficulties in imaging obese patients also lead to delayed diagnoses
of other conditions. For example, increased thickness of compressed
breast tissue results in lower image quality and, when breast cancer is
detected, tumor diameter is significantly larger than in normal weight
patients. In addition, the specificity of mammography decreases
resulting in more false positives and unnecessary further testing. |
|
Scheduling
Before ordering an examination or procedure, check that the patient’s weight does not exceed the relevant weight limit.
If an obese patient is unable to walk and requires transportation by
wheelchair or gurney, it is important to alert the Department of
Radiology to the patient’s needs to enable the staff to have
appropriate transport equipment. Failure to do so leads to delays in
examinations and longer waits for other patients.
Appointments can be scheduled by calling Raul N. Uppot, Editor
,
617-724-9729 or through the Radiology Order Entry system, http://mghroe
.
Further Information
For further questions, please contact , Abdominal Imaging and Intervention at 617-726-8396.
We would like to thank Matthew Hutter, M.D., Department of Surgery, and
Heather Stein, M.D., MGH Weight Center, and James A. Scott, M.D.,
Nuclear Medicine, for their assistance and advice for this issue.
This article provided useful information about the appropriate use of imaging studies:
Note: clicking one of these options will close this window. |
|
|
| |
|
|
References
|
|
|
|
Old, JL, Dusing, RW, Yap, W and Dirks, J. (2005) Imaging for suspected appendicitis. Am Fam Physician 71: 71-8
Oto, A, Ernst, RD, Shah, R, Koroglu, M, et al. (2005) Right-lower-quadrant pain and suspected appendicitis in pregnant women: evaluation with MR imaging--initial experience. Radiology 234: 445-51
Pinto Leite, N, Pereira, JM, Cunha, R, Pinto, P and Sirlin, C. (2005) CT evaluation of appendicitis and its complications: imaging techniques and key diagnostic findings. AJR Am J Roentgenol 185: 406-17
Rao, PM, Rhea, JT, Novelline, RA, Mostafavi, AA and McCabe, CJ. (1998) Effect of computed tomography of the appendix on treatment of patients and use of hospital resources. N Engl J Med 338: 141-6
Rao, PM, Rhea, JT, Rattner, DW, Venus, LG and Novelline, RA. (1999) Introduction of appendiceal CT: impact on negative appendectomy and appendiceal perforation rates. Ann Surg 229: 344-9
Terasawa, T, Blackmore, CC, Bent, S and Kohlwes, RJ. (2004) Systematic review: computed tomography and ultrasonography to detect acute appendicitis in adults and adolescents. Ann Intern Med 141: 537-46
Sivit CJ, Siegel MJ, Applegate, KE, Newman, KD. When Appendicitis Is Suspected in Children RadioGraphics 2001; 21:247–262
|
|
|
|
|
|
|
|